Have an Inquiry

Call (08) 6240 5600

+61 8 6240 5600
+61 8 6240 5600

Shop by Category

HACH DR6000 UV VIS SPECTROPHOTOMETER WITH RFID

SKU HA-LPV441.99.00012 Categories ,

$15,341.70 inc

Available on backorder

Add to Wishlist
Add to Wishlist

Description

The DR6000 is the industry’s most advanced lab spectrophotometer. It offers high-speed wavelength scanning across the UV and Visible Spectrum, and comes with over 250 pre-programmed methods including the most common testing methods used today. With optional accessories allowing for high-volume testing via a carousel sample changer, and increased accuracy with a sample delivery system that eliminates optical difference errors, this instrument ensures you are ready to handle your wide-ranging water testing needs.

Combining the DR6000 with Hach’s unique TNTplus reagent vials, you gain additional accuracy with guided step-by-step testing procedures, while making scratched, flawed or dirty glassware a non-issue due to averaging 10 readings and discarding outliers. The instrument also utilizes RFID technology, allowing it to let you know automatically if your TNTplus reagents have expired, while detecting coefficient factors to avoid errors which can occur in lot-to-lot variations in the chemistries.

Partnering all of these features with an integrated QA software package allows for scheduling, documenting and interpreting your needed quality measurements.

  • Your Water Testing Needs, All in One Spectrophotometer
  • Accessories for High Volume and High Accuracy Testing Needs
  • Advanced Quality Assurance at Your Fingertips
  • Guided Procedures and Elimination of False Readings
  • Automatically Avoids Errors

 

What’s included?
Printed instrument manual, power cords (115V and 230V), universal holder, dust cover, 1-in. square matched glass sample cells, and a universal-adapter

 

Specifications

Data Logger: 5000 data points (result, date, time, sample-ID, user-ID)
Display: TFT 7 inch WVGA color touch
Enclosure Rating: IP20 with closed lid
Interfaces: USB type A (2), USB type B, Ethernet, RFID module
Operating Conditions: 10 – 40 °C, max. 80% relative humidity (non-condensing)
Operating Mode: Transmittance (%), absorbance and concentration (wavelength, time)
Optical System: Reference beam, spectral
Photometric Accuracy: 5 mAbs @ 0.0 – 0.5 Abs, <1% @ 0.5 – 2.0 Abs @ 546 nm
Photometric Linearity: 0.005 – 2 Abs, ≤ 0.01 at > 2 Abs with neutral glass at 546 nm
Photometric Measuring Range: ± 3 Abs
Power Requirements: 100 – 240 V; 50/60 Hz
Preprogrammed Methods: > 240
Sample Cell Compatibility: Rectangular: 10, 20, 30, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch
Optional 100 mm rectangular cell with additional adapter
Scanning Speed: 900 nm/min (in 1 nm steps)
Source Lamp: Tungsten (visible range), deuterium (UV range)
Specific Technology: RFID for easy method update, sample ID and Certificate of Analysis
Spectral Bandwidth: 2 nm
Storage conditions: -25 to 60 °C / max. 80% relative humidity (non-condensing)
Stray Light: KI-solution at 220 nm < 3.3 Abs/ < 0.05%
User Programs: 200
Warranty: 12 months
Wavelength Accuracy: ± 1 nm
Wavelength Range: 190 – 1100 nm
Wavelength Reproducibility: < 0.1 nm
Wavelength Resolution: 0.1 nm
Wavelength Selection: Automatic, based on method selection
Weight: 11 kg

Additional information

Brand